AM8060 Precision Thermometer
User’s Guide
9/43
A “>” will toggle in front of “t1” or ”t2” to indicate which probe
is being measured. Each toggling instance is one measurement.
Line 4, which is indicated with “t1-t2”, shows the temperature
difference of the two probes.
4.4 Selecting Units
Users can select to display ºC, ºF or Ω by scrolling touch keys up
or down (
or
)
4.5 Selecting Sensors/Probes and Adjusting Parameters
4.5.1 Probe Types
AM8060 can work with different types of PRTs. Users can
choose probe types and input the parameters of the probes.
The menu key
MENU
and return/right shift key
can be used
for probe selection and parameter adjustment.
The probes compatible with AM8060 are:
25 Ω or 100 Ω SPRTs, or
25 Ω or 100 Ω Secondary PRTs, or
100 Ω IPRTs.
The temperature coefficient of these probes can either be
0.00385 Ω/Ω/°C or 0.003925 Ω/Ω/°C.
The probe types working with AM8060 are classified and
displayed as IPRT and SPRT: