beautypg.com

AccuMac AM8060 User Manual

Page 10

background image

AM8060 Precision Thermometer

User’s Guide

9/43

A “>” will toggle in front of “t1” or ”t2” to indicate which probe

is being measured. Each toggling instance is one measurement.

Line 4, which is indicated with “t1-t2”, shows the temperature

difference of the two probes.

4.4 Selecting Units

Users can select to display ºC, ºF or Ω by scrolling touch keys up
or down (

or

)

4.5 Selecting Sensors/Probes and Adjusting Parameters

4.5.1 Probe Types

AM8060 can work with different types of PRTs. Users can

choose probe types and input the parameters of the probes.
The menu key

MENU

and return/right shift key

can be used

for probe selection and parameter adjustment.

The probes compatible with AM8060 are:

25 Ω or 100 Ω SPRTs, or

25 Ω or 100 Ω Secondary PRTs, or

100 Ω IPRTs.

The temperature coefficient of these probes can either be

0.00385 Ω/Ω/°C or 0.003925 Ω/Ω/°C.

The probe types working with AM8060 are classified and

displayed as IPRT and SPRT: