AM8060 Precision Thermometer
User’s Guide
22/43
resistance is calculated, and the following message will
appear:
25.9977
Rt =25.0000
OHM
Menu= Exit Enter = Yes
25R Rs =
OHM
* 25 Ohm CALIBRATION
5.2.6 The value on the third line is the parameter to be adjusted and
this value must match the standard resistance of the 25Ω
resistor. For example, the above figure shows the current
value as 25.9977, but the standard resistor’s value is 24.9977.
This value can be adjusted by using
or
key to move
the “_” (underscore) to the desired, and using
or
key
to adjust the value. After the adjustment, the instrument will
display:
24.9977
Rt =24.9977
OHM
Menu= Exit Enter = Yes
25R Rs =
OHM
* 25 Ohm CALIBRATION
5.2.7 It is required to wait for 10 seconds to ensure the reading is
stable. Each second is shown with a blinking dot at the top
left hand corner of the display. The next step is to save the
calibrated data to the instrument. To start the data saving,