AccuMac AM8060 User Manual
Page 16

AM8060 Precision Thermometer
User’s Guide
15/43
value or press
MENU
key to exit the program without
saving and go back to the parameter adjustment menu.
4.5.2.8
The following is a4 adjustment procedure
(Continuing from step 4.5.2.4). Pressing
and
keys
to move the cursor to “>SET a4”, and then press
key
to go into the a4 adjustment program, as seen below:
The second line (a4 = -4.129934e-05) displays the current
a4 value, while the next two lines show the adjustment
selections. The a4 adjustment is different from Rtp, as the
value of a4 can be “0”. If a4 is “0”, the user can move the
cursor by using
and
keys to “SET A4=0” line and
press the
key to confirm the selection. If a4 is a value
other than “0”, the user can move the cursor to “ADJUST
a4” line and then press
key to go into the a4
adjustment program.