AccuMac AM8060 User Manual
Page 20

AM8060 Precision Thermometer
User’s Guide
19/43
* 100 Ohm CALIBRATION
Connect a 100 Ohm
Standard Resistor
and Press Enter
Menu= Exit Enter = Yes
-------------------------------------------
5.1.4 The instrument will show the phrase “Connect a 100 Ohm
Standard Resistor and Press Enter” to indicate that a 100Ω
must be connected to sensor1 (S1) terminals as Figure3.
Users can exit the calibration by pressing
MENU
key to go back
to the main menu (*CALIBRATION MENU*).
5.1.5 After the probe is correctly connected, press
key to start
the calibration. The instrument will display “Waiting…”
while the resistance is calibrated. When the process is
finished, the screen will display the result as seen below:
* 100 Ohm CALIBRATION
100.0238
Rt = 111.0000
OHM
Menu= Exit
<^v>=Adj
100R Rs =
OHM
5.1.6 To calibrate the instrument, users must adjust the value
(100.0238 as shown above) to match the standard resistor’s
resistance. For example, if the resistance of the standard
resistor is 100.0038 ohm, users can move the underline
cursor to the desired position using
and
keys, and