2 specifications – AccuMac AM8060 User Manual
Page 6

AM8060 Precision Thermometer
User’s Guide
5/43
2 Specifications
2.1 Specifications
Temperature range:-200 ºC ~ 850 ºC
Resolution:0.001 ºC
Accuracy:
25Ω standard platinum
temperature probe
100Ω standard or industrial
platinum temperature probe
±0.01 ºC @ -196 ºC
±0.010 ºC @ -200 ºC
±0.008 ºC @ 0 ºC
±0.008 ºC @ 0 ºC
±0.009 ºC @ 232 ºC
±0.009 ºC @ 232 ºC
±0.015 ºC @ 420 ºC
±0.015 ºC @ 420 ºC
±0.025 ºC @ 660 ºC
±0.020 ºC @ 660 ºC
±0.025 ºC @ 850 ºC
Short term stability:
o ±0.001 ºC(with 100Ω platinum probe)
o ±0.003 ºC(with 25Ω platinum probe)
Working environment:15ºC ~ 35 ºC
Power supply:AC 100V ~ 220V
Power consumption:< 15 W
Dimensions :180 mm x 65 mm x 200 mm
Weight :0.7 lbs (300 g)
2.2 Probe Types
SPRT or PRT with temperature coefficient α=0.003925
(W
Ga
≥1.11807).
IPRT with temperature coefficient α=0.00385.