AccuMac AM8060 User Manual
Page 11

AM8060 Precision Thermometer
User’s Guide
10/43
If the probe is an industrial or precision platinum
resistance thermometer probe with a temperature
coefficient of 0.00385 Ω/Ω/°C, AM8060 will categorize
the probe as “IPRT”. AM8060 calculates temperature
according to CVD (IEC751) equation.
If the probe is a precision platinum resistance
thermometer probe or secondary PRT with temperature
coefficient of 0.003925 Ω/Ω/°C (W
Ga
>=1.11807),
AM8060 will categorize the probe as “SPRT”. AM8060
calculates temperature according to ITS-90 standard
platinum resistance thermometer equations.
If the probe is a standard platinum resistance thermometer,
the probe will be displayed as “SPRT”. AM8060
calculates temperature according to ITS-90 standard
platinum resistance thermometer equations.
4.5.2 Selecting Sensors/Probe and Parameters Input
4.5.2.1
To select the type of sensor, press
MENU
key and then
hold
key. After 3 seconds, AM8060 will display the
parameter adjustment main menu:
There are four options for parameter adjustments: