5 setting the ethernet communication parameters, See section 2.5 – LumaSense Technologies INNOVA 1412i User Manual
Page 19

Chapter 2
______________________________________________________________________
_____________________________________________________________________________
BE6025-15
1412i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 19 of 202
2.4.4
Checking/Changing the Data Log and Error Log
Options.
The Data Log and Error Log must be set using the push-buttons on
the front of the monitor.
Press SET-UP S3 S1 S1 S3 ▼S1 S1
Press SET-UP to exit the set-up mode
2.5 Setting the Ethernet Communication Parameters
The LumaSoft Gas 7810 application with Ethernet communication is
able to communicate with the 1412i gas monitor through a local
Ethernet network.
The homepage of the 1412i gas monitor can also be displayed in a
web browser, when the 1412i gas monitor is connected to a local
Ethernet network.
2.5.1
Checking/Changing the Ethernet Communication
Parameters
The communication parameters for the Ethernet interface can be set
by using the push-buttons on the front of the monitor.
1. Press SET-UP S3 S1 S3 S2. The screen display now shows the
following text.
2. Press S1 to enter the Ethernet setup. During this setup you can
also use the ▲ and ▼ keys to go back and forth between the
available Ethernet settings. The screen display now shows the
following text.
3. Here you can select whether the IP address of the 1412i gas
monitor is to be assigned by a DHCP server in the local network.
If you want the 1412i's IP address to be assigned by the DHCP
server press the S3 (Yes) key, where after you will proceed to the
SELECT ADDRESSABLE SET-UP BRANCH
Ethernet IEEE 488
IS 1412i'S IP ADDRESS SET BY DHCP ?
NO
YES