periodically taken out and cleaned. We recommend that this task is
done at least as often as calibration, and more frequently if the
Monitor is working in a very dusty atmosphere or an atmosphere
containing high concentrations of other particulate matter.
Use a “pozidrive” screwdriver to unscrew the screw on the grid cover.
Remove the filter pad from inside the grid cover. Wash the filter pad
in warm water containing some liquid soap. Rinse all the soap out of
the filter, by running clean water through it, and let it dry
thoroughly.
Place the clean, thoroughly dry, filter pad inside the grid cover.
Screw the grid cover firmly back in place over the ventilation unit.
Fig.11.9 Back panel of the 1412i showing the ventilation unit
11.3
Cleaning the Instrument
It is recommended to clean the instrument using a damped cloth.