6.4 Setting of auto calibration
6.4.1 Auto calibration
Auto calibration is automatically carried out
when zero calibration and span calibration are
set.
Before changing the setting of auto calibration,
set the ON/OFF to “OFF”.
1. During measurement, press the
MODE
key
to display the User mode.
2. Point the cursor to “Setting of Auto
Calibration” by pressing the
or
key. Press the
ENT
key.
3. In the “Setting of Auto Calibration”
screen that appears, point the cursor to
any item you want by pressing the
or
key. Press the
ENT
key.
4. In the “Auto Calibration Parameter
Setting” screen that appears, perform the
value entry or the setting. For the value
entry or setting change, use the
or
key. To change the setting, use the
key to move the cursor to the right.
After setting, press the
ENT
key, and auto
calibration is carried out by the entered
setting value.
(
)
ENT
MODE
(
)
ENT
ENT
End of Auto Calibration Setting
Description of setting items
• Start Time : Setting at the first calibration
(day of the week, hour, minute)
• Cycle
: A period between the start time of one
calibration and another
(unit : hour/day)
• Flow Time : The time required for the calibration gas
to be replaced in the cell
• ON/OFF : Auto zero calibration ON or OFF
To close the "Setting of Auto calibration" or cancel
this mode midway, press the
ESC
key.
A previous screen will return.
To close Setting of Auto calibration
ESC