Setting and calibration, 1 changeover of range – Yokogawa IR200 User Manual
Page 36
6 - 1
IM 11G02M01-01E
6. SETTING AND CALIBRATION
6.1 Changeover of range
This mode is used to select the ranges of measured
components.
1. During measurement, press the
MODE
key to
display the User mode.
2. Point the cursor to “Changeover of Range”.
Press the
ENT
key.
3. The “Channel Selection” screen appears.
Press the “
“ or “
” key until the “ ”
cursor moved selects a desired CH (compo-
nent).
4. After selection, press the
ENT
key.
Note) The range of O
2
correction instanta-
neous values and O
2
correction aver-
age values is automatically switched by
changing the range of instantaneous
value of each CH (component).
5. In the Range Setting screen that appears,
move the cursor by pressing the
or
key to select the range. (The range with a
mark of is currently selected.)
6. After selection, press the
ENT
key.
7. Measurement is conducted within the selected
range. The range identification signal (CO) is
shorted with the low range (Range 1), and
open with the high range (Range 2).
Note) If the Remote Range is set to ON, the
changeover of range cannot be per-
formed on the screen.
To close Changeover of range, or cancel the
commad midway, press the key.
A previous screen will return.
To close Changeover of range
ESC
MODE
(
)
ENT
ENT
(
)
ENT
End of Range Changeover
▲