6 - 9

IM 11G02M01-01E

5. In the “Calibration Range Selection” screen

that appears, select “both” or “current” by
pressing the

or

key.

• When selecting “both”, Range 1 and Range

2 of the set CH are calibrated together.

• When selecting “current”, the range alone

displayed at the set CH is calibrated.

In case of both setting

Two cursors will appear in both ranges

Manual Calibration screen

(

)

ENT

To close Setting of Calibration Range or

to cancel this mode midway, press the ESC key.

A previous screen will return.

To close Setting of Calibration Range

ESC

Note) When calibration is performed by the “both” setting under the normal operating

condition, prepare a span gas cylinder on the normal operating range side. It is
recommend to perform span gas calibration in the normal operating range.

The other range that is calibrated by “both” may result in some error (max. of

±

5% of

FS).

End of Calibtation Range Setting

CH1
CO

2

CH2
CO

Range 1: 0 to 10 vol%

Range 2: 0 to 20 vol%

Range 1: 0 to 500 ppm
Range 2: 0 to 2000 ppm

both

current

CH1: Range 1 and Range 2 are calibrated together, with zero or span calibration.

CH2: Only currently displayed range is calibrated, with zero or span calibration.

Example