2 installation of analyzer, 3 piping, Caution – Yokogawa IR200 User Manual
Page 15
3 - 2
IM 11G02M01-01E
CAUTION
3.3 Piping
Caution on piping
In piping, the following precautions should be observed.
Wrong piping may cause gas leakage.
If the leaking gas contains a toxic component, there is a risk of
serious accident being induced.
Also, if combustible gas is contained, there is a danger of
explosion, fire or the like occurring.
• Connect pipes correctly referring to the instruction manual.
• Exhaust should be led outdoors so that it will not remain in
the locker and installation room.
• Exhaust from the analyzer should be relieved in the atmo-
spheric air in order that an unnecessary pressure will not be
applied to the analyzer. Otherwise, any pipe in the analyzer
may be disconnected to cause gas leakage.
• For piping, use a pipe and a pressure reducing valve to which
oil and grease are not adhering. If such a material is adhering,
a fire or the like accident may be caused.
3.2 Installation of analyzer
There are two methods of installing the analyzer.
For detailed dimensions, see Chapter 9.3.
Types
Guide rail
Slide rail
External dimensions
483
Mounting dimensions
Mounting method
(Unit:mm)
INFRARED GAS ANALYZER
SPAN
ZERO
ENT
ESC
POWER
MODE
429
177
465
M6
Guide rail
101.6
Slide rail
19 inch rack mounting (Note)
450 or more
Note) The mounting method should be selected to meet the installation requirements since the top
cover must be detached from the gas analyzer for maintenance and check.
Mounting method
Slide rail
No maintenance space is provided at the top.
Guide rail
Maintenance space is provided at the top.
Conditions
Recommended slide rail: 305A-20, Accuride International Inc.
Remarks
These methods must be rigid enough to
withstand the mass (about 10 kg) of the gas
analyzer.