2 calibration setting, 1 setting of calibration concentration – Yokogawa IR200 User Manual
Page 37
6 - 2
IM 11G02M01-01E
6.2 Calibration setting
This mode is used to set calibration concentration and actions. The calibration setting involves
calibration concentration, calibration gas, zero calibration, calibration range and auto-calibration
component.
“Calibration Gas” is displayed if NO or SO
2
is included in components to be analyzed.
6.2.1 Setting of calibration concentration
It allows you to set concentrations of the standard gas (zero and span) of each channel used for
calibration.
1. During measurement, press the
MODE
key
to display the User mode.
2. Point the cursor to “Setting about Cali-
bration” by pressing the
or
key.
Press the
ENT
key.
3. In the “Setting about Calibration” screen
that appears, point the cursor to “Calibra-
tion Value” by pressing the
or
key. Press the
ENT
key.
4. In the “Calibration Concentration CH
Selection” screen that appears, point the
cursor to CH you want to set by using the
or
key. Press the
ENT
key.
MODE
(
)
ENT
(
)
ENT
(
)
ENT