3 setting of manual zero calibration – Yokogawa IR200 User Manual
Page 41
6 - 6
IM 11G02M01-01E
1. During measurement, press the
MODE
key
to display the User mode.
2. Point the cursor to “Setting about Cali-
bration” by pressing the
or
key.
Press the
ENT
key.
3. In the “Setting about Calibration” screen
that appears, point the cursor to “About
ZERO Calibration” by pressing the
or
key. Press the
ENT
key.
4. In the “CH Selection” screen that
appears, point the cursor to CH you want
to set by using the
or
key. Press
the
ENT
key.
(
)
ENT
(
)
ENT
MODE
(
)
ENT
6.2.3 Setting of manual zero calibration
If zero calibration is to be made manually, select whether to calibrate all components at once or
each of them separately upon selection.