2 setting of calibration gas – Yokogawa IR200 User Manual
Page 39
6 - 4
IM 11G02M01-01E
(1) Press the
MODE
key in measurement state to
display the User mode.
(2) Move the cursor to Setting about
Calibration using the
or the
key,
and then press the
ENT
key.
(3) On the Setting about Calibration screen
that appears, move the cursor to “Calibra-
tion Gas” using the
or the
key,
and then press the
ENT
key.
(4) On the “Gas Selection” screen that
appears, move the cursor to the gas to be
selected using the
or the
key,
and then press the
ENT
key.
(
)
ENT
(
)
ENT
MODE
6.2.2 Setting of calibration gas
Whether moisture is contained in zero gas and span gas used for calibration can be set in advance
as follows. Make the setting according to the sampling system created.
If sampling system controls the moisture content in the calibration gas (ZERO or SPAN) to
become the same as that in the sample gas to be fed to the analyzer using an electronic cooler,
etc., set “With” water vapor.
(
)
ENT