Device status, Device status -1, 7 . device status – Yokogawa EJA115 User Manual
Page 34
IM 01C22T02-01E
7-1
7. DEVICE STATUS
7 .
DEVICE STATUS
Device setting status and failures of EJA are indicated by using parameter DEVICE_STATUS_1,
DEVICE_STATUS_2 and DEVICE_STATUS_3 (index 1045, 1046 and 1047) in Resource Block.
Table 7.1 Contents of DEVICE_STATUS_1 (index 1045)
Display
through DD
Description
Hexadecimal
Sim.enable
Jmpr On
RB in O/S mode
(AL.21)
AMP Module
Failure 2 (AL.03)
LINK OBJ. 1/17
not open
LINK OBJ. 2
not open
LINK OBJ. 3
not open
LINK OBJ. 4
not open
LINK OBJ. 5
not open
LINK OBJ. 6
not open
LINK OBJ. 7
not open
LINK OBJ. 8
not open
LINK OBJ. 9
not open
LINK OBJ. 10
not open
LINK OBJ. 11
not open
LINK OBJ. 12
not open
LINK OBJ. 13
not open
LINK OBJ. 14
not open
LINK OBJ. 15
not open
LINK OBJ. 16
not open
SIMULATE_ENABLE switch
is ON.
Resource Block is in O/S
mode.
AMP module failure
Link object 1 is not open.
Link object 2 is not open.
Link object 3 is not open.
Link object 4 is not open.
Link object 5 is not open.
Link object 6 is not open.
Link object 7 is not open.
Link object 8 is not open.
Link object 9 is not open.
Link object 10 is not open.
Link object 11 is not open.
Not used for EJA.
Not used for EJA.
Not used for EJA.
Not used for EJA.
Not used for EJA.
0x80000000
0x40000000
0x20000000
0x10000000
0x08000000
0x04000000
0x02000000
0x01000000
0x00800000
0x00400000
0x00200000
0x00100000
0x00080000
0x00040000
0x00020000
0x00010000
0x00008000
0x00004000
0x00002000
0x00001000
0x00000800
0x00000400
0x00000200
0x00000100
0x00000080
0x00000040
0x00000020
0x00000010
0x00000008
0x00000004
0x00000002
0x00000001
T0701.EPS
Table 7.2 Contents of DEVICE_STATUS_2 (index 1046)
Display
through DD
Description
Hexadecimal
Data is out of
LCD display
range (AL.61)
Zero Adjust value
is out of normal
range (AL.64)
Temperalure is
out of normal
range (AL.43)
Static Pressure is
out of normal
range (AL.42)
Differential
Pressure is out of
normal range
(AL.41)
AMP Module
Failure 3 (AL.02)
AMP Module
Failure 2 (AL.03)
AMP Module
Failure 1 (AL.02)
Capsule Module
Failure 3 (AL.01)
Capsule Module
Failure 2 (AL.01)
Capsule Module
Failure 1 (AL.01)
Data is out of LCD display
range.
Zero adjustment value is out
of range.
Alarm is generated when the
value is out of LRL -10% to
URL +10%.
Process temperature is out
of limit.
Alarm is generated when the
temperature is out of -50 to
130 degC.
Static pressure is out of
MWP.
Alarm is generated when the
static pressure exceeds
110% of MWP.
Differential pressure is out of
range.
Alarm is generated when the
value is out of LRL -10% to
URL +10%.
Amp module failure
Amp module failure
Amp module failure
Capsule module failure
Capsule module failure
Capsule module failure
0x80000000
0x40000000
0x20000000
0x10000000
0x08000000
0x04000000
0x02000000
0x01000000
0x00800000
0x00400000
0x00200000
0x00100000
0x00080000
0x00040000
0x00020000
0x00010000
0x00008000
0x00004000
0x00002000
0x00001000
0x00000800
0x00000400
0x00000200
0x00000100
0x00000080
0x00000040
0x00000020
0x00000010
0x00000008
0x00000004
0x00000002
0x00000001
T0702.EPS