IRM-5000P USER’S MANUAL

REV 2

34

3.1.5. Performing a Dielectric Discharge (DD) Test

The dielectric discharge (DD) test measures the dielectric absorption of a tested material. A
typical DD test is timed for 30 minutes. Follow the steps below to perform a dielectric discharge
test:

a. Start from the “START-UP” menu:

Press the

[1]

key (RUN TEST).

b. The following screen will be displayed:

Press the

[3]

key (DD).

c. The following screen will be displayed:

Press the

[1]

key (30 Min).

d. The following screen will be displayed:

Press the

[1]

key to select 500V or press the

[2]

key to enter a custom test voltage

value.

1. 500V
2. Custom

TERMINAL VTG: 1 V

1. 30 Min (Default)

2. Custom

TERMINAL VTG: 1 V

1. R 2. PI
3. DD 4. SV

TERMINAL VTG: 1 V

1. RUN TEST 05/05/10

2.

UTILITIES 10:15:15

Bat:

TERMINAL VTG: 1 V