IRM-5000P USER’S MANUAL

REV 2

16

5. Custom

Press the

[5]

key to enter a custom test voltage value. The following screen will

be displayed:

Type the desired test voltage (between 50 V – 5000 V) using the numeric keypad,
and then press the

[ENTER]

key. Continue to step e.

e. The following screen will be displayed:

1. YES

Press the

[1]

key (YES) if you would like to enter the Device Under Test (D.U.T.)

temperature. The following screen will be displayed:

Type the D.U.T. temperature using the numeric keypad. The screen will be
updated as shown:

Press the

[ENTER]

key and continue to step f.

2. NO

Press the

[2]

key (NO) if you do not want to enter the Device Under Test (D.U.T.)

temperature. Continue to step f.

Test Voltage (50-5K)
0 Volts

TERMINAL VTG: 1 V

ENTER TEMP (0-254 F)
120

°F


TERMINAL VTG: 1 V

ENTER TEMP (0-254 F)

°F


TERMINAL VTG: 1 V

ENTER D.U.T. TEMP?
1. YES 2. NO

TERMINAL VTG: 1 V