REV 2 IRM-5000P USER’S MANUAL

25

The LCD screen will be dynamically updated displaying the remaining time along with
the resistance, current, and test voltage values. When the test is completed, the results
will be displayed as shown:

Press any key to continue.

NOTE

If you had entered a D.U.T. temperature in step e, the IRM-5000P will calculate
the equivalent resistance values at 40

°C and 20°C and display them on the LCD

screen.

i. The following screen will be displayed:

Press the

[1]

key (YES) to print the test report (See Figure 7 for a sample test report

printout. See Table 4 for an interpretation of).

j. The test report will be printed on the built-in thermal printer, and then the following

screen will be displayed:

Press the

[1]

key (YES) to save the test results.

k. The following screen will be displayed:

Press any key to continue.

TEST NUMBER 8
HAS BEEN SAVED!

TERMINAL VTG: 1 V

SAVE TEST?

1.YES 2.NO

TERMINAL VTG: 1 V

PRINT REPORT?

1.YES 2.NO

TERMINAL VTG: 1 V

Res= 40.81MK PI=7.7
350V 99.37uF 10:00

TERMINAL VTG: 3 V