IRM-5000P USER’S MANUAL

REV 2

18

The LCD screen will be dynamically updated displaying the remaining time along with
the resistance, current, and test voltage values. When the test is completed, the results
will be displayed as shown:

Press any key to continue.

NOTE

If you had entered a D.U.T. temperature in step e, the following screen will be
displayed:

This screen shows the equivalent resistance value at 40

°C. The equivalent

resistance value is automatically calculated by the IRM-5000P based on the
D.U.T. temperature entered in step e.

Press any key to continue. The following screen will be displayed:

This screen shows the calculated equivalent resistance value at 20

°C. Press any

key to continue.

i. The following screen will be displayed:

Press the

[1]

key (YES) to print the test report (See Figure 4 for a sample test report

printout).

j. The test report will be printed on the built-in thermal printer, and then the following

screen will be displayed:

D.U.T. TEMP: 200

°F

EQ R @ 20

°C: 1.599T

TERMINAL VTG: 1 V

D.U.T. TEMP: 200

°F

EQ R @ 40

°C: 399.6G


TERMINAL VTG: 1 V

PRINT REPORT?

1.YES 2.NO

TERMINAL VTG: 1 V

Res=989.7M 4.87uA
5000V 0.0uF 10:00

TERMINAL VTG: 1 V