3.1.3. Performing a Polarization Index Ratio (PI) Test

The polarization index (PI) is derived from the ratio of the insulation resistance at ten minutes
divided by the insulation value at one minute. For a typical insulation, the resistance value
increases as time increases during a test.

Standard voltages of 500 V, 1000 V, or 5000 V can be used, or the user can enter a custom test
voltage value from 50 V to 5000 V. The default test duration is ten minutes, but a custom test
duration can be set from 10 to 99 minutes. Follow the steps below to perform a PI test:

a. Start from the “Start-Up” menu:

Press the

[1]

key (RUN TEST).

b. The following screen will be displayed:

Press the

[2]

key (PI).

c. The following screen will be displayed:

1. 10 Min (Default)

Press the

[1]

key to select the default test duration of 10 minutes. Continue to

step d.

2. Custom

Press the

[2]

key to enter a custom test duration (between 10 to 99 minutes).

The following screen will be displayed:

Press the

[ENTER]

key to continue.