3.3.5. Performing a Current Source Test

The EZCT-2000B’s programmable current source can be used to verify CT loads. The EZCT-
2000B can output up to a 20A current (0-15Vac). To perform a current source test:

a. Make connections per the figure below:

b. Start from the “START-UP” menu:

Press the

[5]

key (CURRENT SOURCE).

1.

RUN TEST 06/17/09

2.

SETUP 10:14:25

3.

TEST PLANS

4.

DIAGNOSTICS

5.

CURRENT SOURCE