0 diagnostics, verification, and troubleshooting – Vanguard EZCT-2000B User Manual
Page 101

REV 1 EZCT-2000B USER’S MANUAL
97
5.0 DIAGNOSTICS, VERIFICATION, AND TROUBLESHOOTING
5.1
Performing a Diagnostics Test
The Diagnostics test mode displays the output voltage at the selected X leads (V
x
), the voltage
sensed by the H leads (V
h
), and the X voltage excitation current (I
x
). These values can then be
verified using an external volt meter and ampere meter.
To perform a diagnostic test:
a. Start from the “START-UP” menu:
Press the
[4]
key (DIAGNOSTICS).
b. The following screen will be displayed:
Select the tap connection by pressing the corresponding key (
[1]
–
[4]
). If the tap
connection is not listed, press the
[5]
key to view the next page of options and then
select the correct tap connection.
c. The following screen will be displayed:
Select the maximum output voltage on the X terminals by pressing the corresponding
key (
[1]
–
[5]
).
select voltage range
1.
50v
2.
300v
3.
500v
4.
1200v
5.
2000v
select tap
1.
x1-x2
2.
x1-x3
3.
x1-x4
4.
x1-x5
5.
next page
1.
RUN TEST 06/17/09
2.
SETUP 10:14:25
3.
TEST PLANS
4.
DIAGNOSTICS
5.
CURRENT SOURCE