Vanguard CT-7000 S2 User Manual
Page 93
REV 2 CT-7000 S2 USER’S MANUAL
88
4.4
Printing Raw Test Record Data
You can print the raw hexadecimal timing data collected by the CT-7000 S2 using the “PRINT
DATA” command. This feature is used mainly for factory diagnostic testing.
The CT-7000 S2 collects 20,000 data points in the one-second timing window. As a result, each
data sample is 0.05 ms apart. By looking at the time marker and the data point, you can find the
channel activities vs. time in the timing record.
To print the CT-7000 S2 timing data from memory:
a. Start from the “START-UP” menu:
Press the
[4]
key (DIAGNOSTICS).
b. The following screen will be displayed:
Press the
[4]
key (NEXT PAGE).
c. The following screen will be displayed:
Press the
[1]
key (PRINT DATA).
d. The following screen will be displayed:
Enter the starting address boundary by typing a number from 0 to 9. Pressing the
[0]
key will start printing the data at address location 0, or the first data point. Pressing the
[1]
key will start printing the data at address location .10000 second.
STARTING ADDRESS?:
1. PRINT DATA
2. ENCODER FILTER
3. CONTACT FILTER
1. SLOW CLOSE TEST
2. CHECK HOOKUP
3. TEST TRANSDUCER
4. NEXT PAGE
1. TIME BRKR 02/14/11
2. GET RSLT 10:05:10
3. SETUP
4. DIAGNOSTICS