Gentec-EO PC-Gentec-EO (for MAESTRO and INTEGRA) User Manual
Page 61

PC-Gentec-
EO User’s Manual Revision 5.0
60
:0
0028
E1 20
á
:0
0029
0 3A
:
:0
002A
39 31
1 9
Detector serial number (You must convert the
:0
002B
36 39
9 6
hexadecimal values in ASCII characters)
:0
002C
32 37
7 2
:0
002D
0 0
0000 = Null termination character
:0
002E
D70A
0.0200
Trigger Level LSB (between 0.001 and 0.999)
:0
002F
3CA3
Trigger Level MSB (between 0.001 and 0.999)
:0
0030
0001
1
Is autoscale mode on? LSB
:0
0031
0000
0
Is autoscale mode on? MSB
:0
0032
0000
0
Is anticipation on? LSB
:0
0033
0000
0
Is anticipation on? MSB
:0
0034
0000
0
Is zero offset on? LSB
:0
0035
0000
0
Is zero offset on? MSB
:0
0036
0000
1.0000
Correction Multiplier LSB
:0
0037
3F80
Correction Multiplier MSB
:0
0038
0000
0.0000
Correction Offset LSB
:0
0039
0000
Correction Offset MSB
:1
0000
0000
0
End of structure
Noise suppression (Integra only)
Sets or Queries the sampling size of the noise suppression. For joulemeter and UM only.
Text Command
Parameters
Return
AVG###
###: average size
none
Example
Response from Monitor
Send avg016 sets sampling size to 16 pulses, Integra replies
Ok.
Valid averages are powers of 2 from 1 to 128. If a non valid
argument is sent, An Error is returned and the value is left
unchanged.
Get current value with frequency (Integra only)
The *CTU can be called send current value with frequency.