Text Command
Parameters
Return
GRR
None
Data in ASCII
Set the joulemeter binary mode
This command is used to set the monitor in binary or ASCII mode. Refer to section 3.4.2 for the
MAESTRO binary mode description.
Command
Parameters
Return
SS1
0= ASCII
1= Binary
Get Binary Joulemeter Mode
This command returns whether or not the binary joulemeter mode is activated for serial
communication. Refer to section 3.4.2 for the MAESTRO binary mode description.
Text Command
Parameters
Return
GBM
None
1: On
0: Off
Example
Response from Monitor
Command:
*GBM
Binary Joulemeter Mode : 0 \r\n
Set Analog Output
This command is used to enable or disable the output of the current value on the analog port of
the device.
Text Command
Parameters
Return
ANO
1 to Enable, 0 to Disable
MEASUREMENT COMMANDS -- SETUP
Set Personal Wavelength Correction
This command is used to specify the wavelength in nm being used on the detector. The
EEPROM in the detector contains measured spectral data for a wide range of wavelengths. If the
wavelength input by the user is different from the predefined list of wavelengths on the device, a
custom value is interpolated. Specifying zero as a wavelength or providing an out-of-bound value
as a parameter restores the default settings. A valid value is set between the lowest and highest
wavelengths supported by the device, and it should not be a floating point value. The input
parameter must have 5 digits. If the desired wavelength does not have 5 digits you must enter a
zero-padded number. For example 514nm, you must enter 00514.