Gentec-EO PC-Gentec-EO (for MAESTRO and INTEGRA) User Manual
Page 49

PC-Gentec-
EO User’s Manual Revision 5.0
48
The value should be set between 0.1 and 99.9.
Example
Response from Monitor
Text Command:
*STL15.4 (15.4%)
*STL0.20 (.2%)
Get Trigger Level
This command returns the trigger level in %. The value is between 0.1% and 99.9%.
Text Command
Parameters
Return
GTL
None
Returns the trigger level
in %.
Example
Response from Monitor
Command:
*GTL
Trigger Level : 2.0 \r\n
Get Mode Display
This command ret
urns the MAESTRO’s power mode. Depeding on the head, it can be power
mode in W, power mode in dBm (DBM), energy mode in J, Single Shot Energy mode in J (SSE),
or simply no detector is connected to the MAESTRO.
Text Command
Parameters
Return
GMD
None
POWER = 0
ENERGY = 1
SSE = 2
DBM = 6
NO_DETECTOR = 7
Example
Response from Monitor
Command:
*GMD
Mode : 0 \r\n
MEASUREMENT COMMANDS -- DATA ACQUISITION
Query Current Value
This command is used to query the value that is currently being displayed on t
he device’s screen.
The value is displayed in watts or in joules.
Text Command
Parameters
Return
CVU
None
Current value