Tips
If
“Save statistics to the acquisition file” is checked, the Continuous mode
cannot be used. The statistics are entered into a file at the end of the
duration period, so we must use the Fixed duration.
The file used for the statistics is the same as the file used for the
acquisition. Therefore, if the filename of the statistics is changed, the
acquisition log filename will be changed automatically.
To store only the statistics in a file, check the
“Save statistics to the
acquisition file
” and start the statistics. To add statistics at the end of an
acquisition log file, check the
“Save statistics to the acquisition file” option
and start the acquisition. The statistics will automatically start at the same
time and will be entered in the file at the end of the acquisition.
Statistical Parameters
Power
Energy Definition
Current value
Value of the most recent measurement
Average value
Average from the start of values in the sample, E
avg
or P
avg
Maximum value
Highest value in the sample period, E
max
or P
max
Minimum value
Lowest value in the sample period, E
min
or P
min
RMS stability
Root mean square stability represents the standard deviation as
a percentage of the average.
100
avg
E
STD
RMS
,
100
avg
P
STD
RMS
PTP Stability
Shows the spread between the highest and lowest point in the
sample as a percent.
100
min
max
avg
E
E
E
PTP
,
100
min
max
avg
P
P
P
PTP
STD Deviation
(only available for the
single full-screen mode)
A measure of the spread of the data around the average.
1
1
2
n
E
E
STD
n
i
avg
i
,
1
1
2
n
P
P
STD
n
i
avg
i
Repetition Rate
(only available for the
single full-screen mode)
Frequency of pulses coming from the laser, PRR
Average Power
(only available for the
single full-screen mode)
Power calculated from the pulse energies and repetition rate.
PRR
E
P
avg
avg