Maintenance

Model 6700C TOC

Teledyne Analytical Instruments

54

Procedure:

CAUTION: OBSERVE NORMAL PRECAUTIONS WHEN

HANDLING ACIDS.


Replacement

1.

Turn machine OFF.

2.

Turn carrier gas (Air/O

2

) OFF.

3.

Remove and drain tubing and Sparger.

4.

Remove Sparger from spring clips.

Cleaning

1.

Run warm water through tubing to flush out any solid
contamination.

2.

Reinstall Sparger in reverse order.

5.3.10 Injector Assembly

To remove the Injector Assembly (See Figure 5-9):

1. Turn off power

2. Disconnect all tubing

3. Remove Injector

4. Reinstall new injector in reverse order