Maintenance

Model 6700C TOC

Teledyne Analytical Instruments

46

 *Remove ‘O’-Ring and

Sapphire window taking care
to avoid scratching sapphire
window.

*Recommend using toothpick.

Clean sapphire windows with
a soft, lintless tissue (use DI
water, if necessary).

To Reassemble:

 Install IR Cell.

 Install Source Assembly end

piece.

Note: Large outlet port goes
into Detector Assembly (left)

 Insert lock-tight on three thumbnuts to prevent back off.
 Install insulation. Replace assembly as before.
 Recalibrate NDIR (Section 4.4.4).