High Temperature Combustion

Installation/Startup

Teledyne

Analytical

Instruments

27

Figure 3-8: Top Retainer Installation

3.5 Startup of the Analyzer

Before starting the analyzer, please check the following:

Air/oxygen supply – This must provide a constant pressure of

15 ± 2 psi, ultra-pure oxygen or hydrocarbon and CO

2

–free air at

a flow rate up to 300 cc/minute. Pre-purified air may be used
with optional Zero Air Generator which requires Instrument Air
or Oxygen Generator which requires electricity only.

Drain Line – Check to see if the Drain Line is sloped downward

and is free of kinks, loops and allows normal vented gravity flow
with an air break to an open receptacle (jar, pipe, etc.).

Vents – Vent lines must be properly installed and vented to

atmospheric conditions and not allowed to have either a
restriction or a pumping of exhaust flow, as either of these
conditions may result in improper analyzer operation.