To remove and service the NDIR, follow the illustrated sequence as

described in the following NDIR Module Exchange Directions. See
Figure 5-1.

Figure 5-1: Removing the NDIR

Replacement:

1. Turn off power.

2. Remove inlet tube from NDIR Assembly.

3. Remove vent tube from NDIR Assembly.

4. Unplug interface cable from NDIR Assembly.

5. 5. Remove NDIR Assembly from cabinet.

Note: Appearance may differ depending on range of machine.