Maintenance

Model 6700C TOC

Teledyne Analytical Instruments

52

Procedure:

1.

Shut off Power to Instruments.

2.

Shut off all gas to analyzer and bleed system for “0” pressure.

3.

Disconnect fittings/tubing.

4.

Loosen screws holding Module in-place.

5.

Remove Module.

6.

Recalibrate analyzer after re-installation of exchange Module.

Note: Mass Flow Controllers of different manufacture may be

used. Include the manufacturer, make and model
information when communicating with TAI Customer
Service.

5.3.8 Metering Valve

If no O

2

/air flow to the Sparger is observed (no bubbles), check for

tubing and fitting leaks or restrictions. Adjust the Metering Valve being
careful to return to previous indicated bubble rate. If flow cannot be
restored, replace the Metering Valve according to the procedure given
below. Refer to Figure 5-7.

Figure 5-7: Metering Valve