Operation

Model 6700C TOC

Teledyne Analytical Instruments

32

4.3.2 Edit/Alarms

The next recommended settings to be verified or set are the Alarm

Settings.

From the [Edit] Menu, select [Alarms]. The following Menu

appears:

From this menu you can verify or select the desired Alarm points.

4.4 Cal

Although the instrument has been calibrated to your specifications

at the factory, it should be rechecked after satisfactory installation and
periodically calibrated as suggested for your application.

The analyzer calibration is performed by an “end-to-end” method,

whereby a known chemical standard solution is introduced to the
analyzer and the analyzer is “spanned” to that value.

4.4.1 Organic and Inorganic Carbon Standards

KHP (Potassium Hydrogen Phthalate) is recommended for the

organic carbon standard solution. Sodium Carbonate is recommended
for the inorganic carbon standard. Table 3-1 provides the concentration
to be used for two ranges of different chemical compounds (organic &
inorganic carbon). Using a ratio of these concentrations will provide
other ranges for the span solution.