Maintenance

Model 6700C TOC

Teledyne Analytical Instruments

42

Table 5-1 Troubleshooting

PROBLEMS SOLUTION

Analyzer will not turn on

Check fuse & power connections.

Look for loose or damaged wiring.

Check for 12V DC from power supply.

Check power supply fuse.

Analyzer will not calibrate, is
sluggish

Look for & correct any liquid or gas line
leaks.

Analyzer is excessively noisy

Check for leaks at tubing connection

Display not functioning

Check cable connections

Check power supply voltage

Suspect: Malfunctioning display — replace.

Injector not operating (Syringe
plunger not moving back and
forth)

Refer to section “Injection System”

Verify that carrier loss did not cause the
injector to be turned off.

Verify ribbon cable connection to injector.

Injector operating but not
pumping fluid

Exchange injector (P/N )

Refer to section “Injector”

Change tubing

Verify DI line is in full DI container.

5.3 Module Service

5.3.1 NDIR Calibration

(Refer to Section 4.4.3).

5.3.2 NDIR Service

(P/N CP2805 or CP2802 determined by measurement range)