Teledyne 6700-C Total Organic Carbon Analyzer (Hi-Temp Combustion version) User Manual

Page 49

background image

High Temperature Combustion

Maintenance

Teledyne

Analytical

Instruments

49

6. 6. Recalibrate liquid Zero and Span.

Figure 5-4: Removing Computer Module

5.3.6 Reactor Assembly and Gas/Liquid Separator

(P/N R3332) Reactor
(P/N G608) GLS

If Reactor is malfunctioning or GLS is broken, use the following

procedure for removal. Also, see Section 3.2.2 for information on
repacking the Reactor. See Figure 5-5.

Procedure:

Replacement

1. Turn off power to Reactor.

2. Verify ambient temperature. The Reactor temperature is

displayed at the top center of the display.

CAUTION:

HOT SURFACES! WEAR GLOVES TO AVOID
SEVERE BURNS.