Operation

Model 6700C TOC

Teledyne Analytical Instruments

36

To set IR “Zero”:

 Flow 200cc/minute oxygen (or CO

2

free air) for 5 minutes.

 Allow the NDIR reading to stabilize. Select [ZERO]. The

NDIR Reading will be set to 0.

To set IR “Span”, a gas calibration standard mixture of 1% CO

2

in

pure nitrogen is required for the 6 inch (shorter) IR bench. A gas
mixture of 0.1% CO

2

in pure nitrogen is required for the longer (15 inch)

bench.

 Flow 200 cc/minute of this cal gas.
 Allow the NDIR reading to stabilize and then select [SPAN].

The NDIR reading will be set to 10,000.

 Select [OK] to save or [CANCEL] to reject the calibration

setting.

The NDIR has now been properly calibrated. Re-connect tubing as

before.

4.4.4.“End-to End” Calibration

Note: On the initial Startup, the gas calibration should be

checked and, if necessary, performed prior to liquid
calibration.

From Menu, select [CAL]. The following screen will appear: