Installation/Startup

Model 6700C TOC

Teledyne Analytical Instruments

20

Figure 3-1: Combustion Module Interior

3.2 Non – Catalytic Reactor Packing

The following pages describe the assembly of the Reactor using the

“Non-Catalytic” Oxidation Method at temperature up to 950°C.

Note: The Reactor life is directly related to temperature. It is

recommended that the operating temperature not exceed
900°C and preferably be at 850°C.

3.2.1 Identify Parts

Quartz Sleeve (P/N: S1795)