CAUTION: Be careful to slowly place the assembly in a

vertical position, AVOIDING A RAPID DROP of the
Ceramic Sleeve and Spacer to the bottom of the
Reactor.

If not carefully inverted, the Ceramic Sleeve and
Spacer will damage the bottom of the Reactor by
dropping too fast.

3.3 Catalytic Reactor Packing

The following figures represent the packing used for the catalytic

oxidation method.

The optimum furnace operating temperature is between 680°C and

750°C. The lower the temperature, the longer the life of the catalyst.

Figure 3-4: Catalyst Packing


1. Loosely pack Ceramic Wool at bottom of Reactor base.

2. Pour catalyst into Reactor.

3. Loosely pack Ceramic Wool on top of catalyst.