beautypg.com

Appendix – Teledyne 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range User Manual

Page 81

background image

Paramagnetic O2 Purity Analyzer

Appendix

Teledyne Analytical Instruments

81

Appendix

A-1 Model 3010MA Specifications

Sensor:

Paramagnetic

Range:

3 ranges, user selectable from 0-100% O

2

Alarms:

One system-failure alarm contact to detect

power failure

Two adjustable concentration threshold

alarms with fully programmable setpoints.

Displays:

2 line by 20 character, alphanumeric, VFD

screen
One 5 digit LED display

Digital Interface:

Full duplex RS-232 communications port.

Power:

100-130, or 200-240 VAC, at 47/63 Hz.

(User selected on rear of control unit)

Operating Temperature: 20-30

°C

Accuracy:

±0.045%

O

2

absolute at constant

temperature (once thermal equilibrium has
been achieved)
±0.05% O

2

over operating temperature range

20-30

C (once thermal equilibrium has been

achieved)

Response

Time:

10% - 90% in 14 sec @ 600 SCCM

Analog outputs:

0-1VDC percent-of-range (Standard)

0-1VDC range ID (Standard)
4-20 mA dc—isolated— percent-of-range
4-20 mA dc—isolated— range ID (Special
option)

Dimensions:

17.7 cm high

 24.9 cm wide  30.5 cm

deep (6.96

  8.7  12.2).

Weight:

Analysis:

(7.8Kg)

17.2Lbs

Control

unit:

(8.2 Kg) 18Lbs