Part 1 Control Unit

Model 3010MA

50

Teledyne Analytical Instruments

4.3.6 Version Screen

Move the < > arrow key to More and press Enter. With Version

blinking, press Enter. The screen displays the manufacturer, model, and
software version information.

4.3.7 Display Sensor Output

The raw output of the sensor can be displayed from the Sensor Output

screen. This is useful during initial sensor calibration where the sensor
output is adjusted to correspond to zero and span values at ambient (or
operating) vent pressure. See Section 6.

To view the sensor output, press the System button to start the

System function.

Contrast Auto—Cal 
PSWD Logout More 

Use the < > arrow keys to blink More, then press Enter.

Version Self—Test 
###.##  #####  

The Output of the sensor in volts is the first number shown on the

second line. The second number is a raw output of the pressure sensor in
A to D counts.

4.4 The Zero and Span Functions

The analyzer can be calibrated using a background gas without

Oxygen as zero and Oxygen for span gas.

NOTE: After installation, an initial sensor calibration may be required

before the instrument is calibrated in order for the pressure
compensation circuitry to work properly. See Section 6.6.

Connect the calibration gases to the analyzer according to the

instructions given in Section 3.4.1, Gas Connections, and in Part II of this
manual observing all the prescribed precautions.

Shut off the gas pressure before connecting it to the analyzer, and

be sure to limit the pressure to 15 psig or less when turning it back on.

Readjust the gas pressure into the analyzer until the flowrate (as read

on the Analysis Unit flowmeter) settles between 300-1000 SCCM,
preferably 500 SCCM.