6.4 Sample System

The paramagnetic sensor used in the Model 3010MA is a partial

pressure device and is sensitive to pressure variations both at the point of
entry into the system and the sample exit (atmospheric pressure). For
example, a 1% change in vent pressure will cause a corresponding 1%
variation in the oxygen reading.

For higher accuracy, a pressure compensation circuit is installed to

compensate for vent pressure fluctuations. A 1/8 inch line continuously
samples the pressure at the sensor outlet in the Analysis unit and delivers
it to the pressure compensation circuit within the Control Unit. The raw
output from the sensor is corrected for pressure variation using a baseline
pressure set during instrument calibration. The output is adjusted
electronically based on the pressure differential between the present
reading and the baseline set during the last calibration.

The linear range of the uncorrected sensor output is 0-1.2VDC. For

proper operation, the raw output must be adjusted to 0VDC at 0% O

2

and

1VDC at 100% O

2

. This allows the sensor output to remain linear during

minor atmospheric fluctuations. The raw sensor output can be accessed
form the alphanumeric display by using the System key.

On the inlet side, the Model 3010MA is supplied with an externally

mounted flow restrictor to limit the sample flow to 300-1000 SCCM. The
actual restrictor recommended depends on the users sample pressure. Note
that the restrictor can be replaced with a metering valve for more
flexibility. The standard units are supplied with PN: R2533. Table 6-3
identifies restrictors available for the pressure ranges shown as well as
available metering valves.

Table 6-3: Pressure Ranges of Available Restrictors

PRESSURE RANGE

TAI RESTRICTOR

2-5 psig

TAI P/N R2532

5-15 psig

TAI P/N R2533

10-20 psig

TAI P/N R2534

15-30 psig

TAI P/N R2535

Brass metering valve

TAI P/N V744

SS metering valve

TAI P/N V745