Teledyne 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range User Manual
Page 32
![background image](/manuals/345643/32/background.png)
Part 1 Control Unit
Model 3010MA
32
Teledyne Analytical Instruments
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact
Digital Remote Cal Inputs: Accept 0 V (off) or 24
VDC
(on) inputs
for remote control of calibration. (See Remote Calibration Protocol
below.) See Table 3-3 for pin connections.
Zero: Floating input. 5 to 24 V input across the + and – pins
puts the analyzer into the Zero mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows Zero mode to terminate when done. A
synchronous signal must open and close the external zero valve
appropriately.
Span:
Floating input. 5 to 24 V input across the + and – pins
puts the analyzer into the Span mode. Either side may be
grounded at the source of the signal. 0 to 1 volt across the
terminals allows Span mode to terminate when done. A
synchronous signal must open and close external span valve
appropriately. See Figure 3-5 Remote Probe Connector.
(The –C option internal valves operate automatically.)
Cal
Contact: This relay contact is closed while analyzer is
spanning and/or zeroing. (See Remote Calibration Protocol
below.)
Table 3-3: Remote Calibration Connections
Pin Function
9
+ Remote Zero
11
– Remote Zero
10
+ Remote Span
12
– Remote Span
40
Cal
Contact
41
Cal
Contact
Remote Calibration Protocol: To properly time the Digital Remote
Cal Inputs to the Model 3010MA Analyzer, the customer's controller
must monitor the Cal Relay Contact.