4.3.5 System Self-Diagnostic Test

The Model 3010MA has a built-in self-diagnostic testing routine.

Pre-programmed signals are sent through the power supply, output board
and sensor circuit. The return signal is analyzed, and at the end of the test
the status of each function is displayed on the screen, either as OK or as a
number between 1 and 3. (See System Self Diagnostic Test in Chapter 5
for number code.)

The self diagnostics are run automatically by the analyzer whenever

the instrument is turned on, but the test can also be run by the operator at
will. To initiate a self diagnostic test during operation:

Press the System button to start the System function.

Contrast Auto—Cal 
PSWD Logout More 

Use the < > arrow keys to blink More, then press Enter.

Version Self—Test 
Sensor Output: ###  

Use the < > arrow keys again to move the blinking to the Self–Test

function. The screen will follow the running of the diagnostic.

 RUNNING DIAGNOSTIC 
 Testing Preamp — 83 

During preamp testing there is a countdown in the lower right corner

of the screen. When the testing is complete, the results are displayed.

Power: OK   Analog: OK 
Preamp: 0 

The module is functioning properly if it is followed by OK. A number

indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting
and Table 5-1 for number-code
information. The results screen alternates for a time with:

Press Any Key 
To Continue... 

Then the analyzer returns to the initial System screen.