beautypg.com

Teledyne 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range User Manual

Page 54

background image

Part 1 Control Unit

Model 3010MA

54

Teledyne Analytical Instruments

In latching mode, once the alarm or alarms trigger, they will
remain in the alarm mode even if process conditions revert to
non-alarm conditions. This mode requires an alarm to be
recognized before it can be reset. In the non-latching mode, the
alarm status will terminate when process conditions revert to
non-alarm conditions.

4. Are either of the alarms to be defeated?


The defeat alarm mode is incorporated into the alarm circuit so
that maintenance can be performed under conditions which
would normally activate the alarms.

The defeat function can also be used to reset a latched alarm. (See

procedures, below.)

If you are using password protection, you will need to enter your

password to access the alarm functions. Follow the instructions in Section
4.3.3 to enter your password. Once you have clearance to proceed, enter
the Alarm function.

Press the Alarm button on the front panel to enter the Alarm

function. Make sure that AL–1 is blinking.

AL—1      AL—2   
 Choose Alarm 

Set up alarm 1 by moving the blinking over to AL–1 using the < >

arrow keys. Then press Enter to move to the next screen.

AL—1 98.00  %     HI 
Dft—N  Fs—N  Ltch—N 

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL–1 #### % (oxygen)

Out-of-range direction, HI or LO

• Defeated?

Dft–Y/N (Yes/No)

• Failsafe?

Fs–Y/N (Yes/No)

• Latching?

Ltch–Y/N (Yes/No).

To define the setpoint, use the < > arrow keys to move the blinking

over to AL–1 ####. Then use the

 arrow keys to change the number.

Holding down the key speeds up the incrementing or decrementing.
(Remember, the setpoint units are percent-of-oxygen.)