Paramagnetic O2 Purity Analyzer

Operation

Teledyne Analytical Instruments

57

or

L—0.00   M—95.00 
H—98.00  Mode—FX/HI 

Press Escape to re-enter the Analyze mode using the fixed range.

NOTE: When performing analysis on a fixed range, if the oxygen concentration

rises above 100% O

2

upper limit (default value) the output saturates at

1VDC. If the reading goes below the minimum range value the output
voltage will drop to zeroVDC However, the digital readout and the RS-
232 output continue to read the true value of the oxygen concentration
regardless of the analog output range. Oxygen values below 0% or
above 100% should not be considered valid.

4.7 The Analyze Function

Normally, all of the functions automatically switch back to the

Analyze function when they have completed their assigned operations.
Pressing the Escape button in many cases also switches the analyzer back
to the Analyze function. Alternatively, you can press the Analyze button
at any time to return to analyzing your sample.

4.8 Signal Output

The standard Model 3010MA Paramagnetic Oxygen Purity Analyzer

is equipped with two 0–1VDC analog output terminals accessible on the
back panel (one concentration and one range ID). The standard model also
includes one isolated 4–20 mA dc current outputs for the oxygen
concentration.

See Rear Panel in Chapter 3, Installation, for illustration.

The signal output for concentration is linear over the currently

selected analysis range. For example, if the analyzer is set on range that
was defined as 98–100% O

2

, then the output would be:

Table 4.1:Linear Analog Output

% Oxygen

Voltage Signal

Current Signal

Output (V dc)

Output (mA dc)

98.0 0

4

98.2 0.1

5.6

98.4 0.2

7.2