2. To change between American and European fuses, remove the

single retaining screw, flip Fuse Block over 180 degrees, and
replace screw.

3. Replace fuse as shown in Figure 5-2.

4. Reassemble Housing as shown in Figure 5-1.

American Fuses

European Fuses

Figure 5-2: Installing Fuses

5.3 System Self Diagnostic Test

1. Press the System button to enter the system mode.

2. Use the < > arrow keys to move to More, and press Enter.

3. Use the < > arrow keys to move to Self-Test, and press Enter.

The following failure codes apply:

Table 5-1: Self-Test Failure Codes

Power

0 OK

1

5 V Failure

2

15 V Failure

3 Both

failed

Analog

0 OK

1

D/A A (0-1 V Concentration)

2

D/A B (0-1 V Range ID)

3 Both

Failed