Teledyne 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range User Manual

Page 46

background image

Part 1 Control Unit

Model 3010MA

46

Teledyne Analytical Instruments

4.3.3.1

E

NTERING THE

P

ASSWORD

To install a new password or change a previously installed password, you
must key in and ENTER the old password first. If the default password is
in effect, pressing the ENTER button will enter the default TETAI
password for you.

Press System to enter the System mode.

Contrast Auto—Cal 
PSWD Logout More 

Use the <

> arrow keys to scroll the blinking over to

 

PSWD

, and press

Enter to select the password function. Either the default TETAI password
or AAAAA place holders for an existing password will appear on screen
depending on whether or not a password has been previously installed.

     T E T A I 
Enter PWD 

or

     A A A A A 
Enter PWD 

The screen prompts you to enter the current password. If you are not

using password protection, press Enter to accept TETAI as the default
password. If a password has been previously installed, enter the password
using the <

> arrow keys to scroll back and forth between letters, and the

 arrow keys to change the letters to the proper password. Press Enter to
enter the password.

If the password is accepted, the screen will indicate that the password

restrictions have been removed and you have clearance to proceed.

PSWD Restrictions 
     Removed 

In a few seconds, you will be given the opportunity to change this

password or keep it and go on.

Change Password? 
=Yes   =No 

Press Escape to move on, or proceed as in Changing the Password,

below.