3.3 Rear Panel Connections—Control Unit

Figure 3-4 shows the Model 3010MA rear panel. On this panel are

connectors for interfacing with the Analysis Unit, RS-232 communication,
power cord, line voltage switch, input/output connections via the 50-pin
equipment interface connector and a gas inlet port for the pressure
compensation circuit.

Figure 3-4: Rear Panel of the Model 3010MA Control Unit

3.3.1 Gas Connections—Control Unit

The only gas connection made to the Control Unit is the pressure

compensation line extending between the Analysis unit and the Control
Unit. A 1/8 inch tube fitting is installed on the rear panel for this purpose.

To connect the pressure compensation line:

1. Insert the 1/8 inch tube into the fitting and finger-tighten the

nut where the tube cannot be rotated freely. Tighten the nut an
additional 1/8 turn beyond finger-tight.

2. Use a wrench to hold the fitting body and keep it from turning

while you tighten the nut using another wrench When
assembling new fittings for the first time: Tighten the nut 1 ¼
turns.