Part 1 Control Unit

Model 3010MA

48

Teledyne Analytical Instruments

     A A A A A 
Retype PWD To Verify 

Wait a moment for the entry screen. You will be given clearance to

proceed.

     A A A A A 
 TO Proceed 

Use the arrow keys to retype your password and press Enter when

finished. Your password will be stored in the microprocessor and the
system will immediately switch to the Analyze screen, and you now have
access to all instrument functions.

If all alarms are defeated, the Analyze screen appears as:

95.00 %   Anlz 
Range: 95 — 100 

If an alarm is tripped, the second line will change to show which

alarm it is:

95.00 %   Anlz 
AL—1 

NOTE: If you log off the system using the logout function in the system

menu, you will now be required to re-enter the password to gain
access to Span, Zero, Alarm, and Range functions.

4.3.4 Logout

The Logout function provides a convenient means of leaving the

analyzer in a password protected mode without having to shut the
instrument off. By entering Logout, you effectively log off the instrument
leaving the system protected against use until the password is reentered.
To log out, press the System button to enter the System function.

Contrast Auto—Cal 
PSWD Logout More 

Use the < > arrow keys to position the blinking over the Logout

function, and press Enter to Log out. The screen will display the message:

Protected Until 
Password Reentered