Part 1 Control Unit
Model 3010MA
48
Teledyne Analytical Instruments
A A A A A
Retype PWD To Verify
Wait a moment for the entry screen. You will be given clearance to
proceed.
A A A A A
Use the arrow keys to retype your password and press Enter when
finished. Your password will be stored in the microprocessor and the
system will immediately switch to the Analyze screen, and you now have
access to all instrument functions.
If all alarms are defeated, the Analyze screen appears as:
95.00 % Anlz
Range: 95 — 100
If an alarm is tripped, the second line will change to show which
alarm it is:
95.00 % Anlz
AL—1
NOTE: If you log off the system using the logout function in the system
menu, you will now be required to re-enter the password to gain
access to Span, Zero, Alarm, and Range functions.
4.3.4 Logout
The Logout function provides a convenient means of leaving the
analyzer in a password protected mode without having to shut the
instrument off. By entering Logout, you effectively log off the instrument
leaving the system protected against use until the password is reentered.
To log out, press the System button to enter the System function.
Contrast Auto—Cal
PSWD Logout More
Use the < > arrow keys to position the blinking over the Logout
function, and press Enter to Log out. The screen will display the message:
Protected Until
Password Reentered